logo
×

PRODUCTS

T3 Automated platform for the determination of Physicochemical Properties

  • You Are Here :HOME > PRODUCTS > T3 Automated platform for the determination of Physicochemical Properties >
  • T3 Automated platform for the determination of Physicochemical Properties

    description: Basic Introduction: Automated platform for the determination of Physicochemical Properties Elucidate drug potential Globally renowned as the gold standard technique for determination of pKa, logP/D and solubility of ionisable compounds. Rich insights

    Basic Introduction:

    Automated platform for the determination of Physicochemical Properties

    Elucidate drug potential

    Globally renowned as the gold standard technique for determination of pKa, logP/D and solubility of ionisable compounds.

    Rich insights to discovery scientists allowing them to make accurate and assured decisions about their drug candidates.

    Determine pKa values using two techniques: spectrometric (for UV-active compounds) and potentiometric (for non-UV active compounds). With high throughput automated screening capabilities, SiriusT3 can complete up to 80 assays a day.

    Determine the intrinsic solubility of the different forms (crystalline and amorphous) for polymorphs compounds using the Pion’s CheqSol technique.

    Product Advantages:

    Requiring only sub-milligram sample quantities in most cases, the SiriusT3 can complete the analysis in less than 6 minutes per titration using the high throughput pKa screening**.**

    Determine pKa values using two techniques: spectrometric (for UV-active compounds) and potentiometric (for non-UV active compounds).With high throughput automated screening capabilities, SiriusT3 can complete up to 80 assays a day.‍

    Measure log P/D as a function of pH in less than 2 hours using the potentiometric technique in place of traditional shake-flask methods. Determine kinetic and intrinsic solubility using Pion’s patented CheqSol technique within two hours.

    Use the CheqSol technique to provide intrinsic solubility as a function of pH and generate detailed information about the extent and duration of the supersaturated states.

    Product Key Features:

    DTAu –Fully automated with pKa high throughput screening SiriusT3 system with both spectroscopic and potentiometric assay capabilities.

    DTu – A single vial system for one assay at a time with both spectroscopic and potentiometric assay capabilities.

    DTA – A fully automated system with only potentiometric assay capabilities.

    DT - A single vial system for one assay at a time with only potentiometric assay capabilities.

    Assay Expert – Optional integrated prediction module

    Application:

    Measurement of logP and logD.

    Measurement of extreme pKas.

    Measurement of logP for samples poorly soluble in water and octanol.

    Solubility and supersaturation.

    Determination of solubility - CheqSol explanation.

    Detailed Specifications

    Specification

    pH Electrode: Ag/AgCl, double junction reference
    pH Range: 1.8-12.2, calibrated via Sirius Four-Plus™
    Electrode Storage: Stored in Home position; calibrated with pH 7 buffer
    Stirring: Overhead, variable speed, computer-controlled
    Temperature: Measured at each data point by thermocouple, Peltier controlled, 12℃-70℃
    Turbidity Detection: Light transmission detection - percentage of light transmission (0.1%)
    Reagents: Water, acid, base, co-solvent, dispensing solvent
    Dispensing Solvents: Up to 7 types, including n-octanol, dodecane, toluene
    Co-Solvents: Up to 10 types, including methanol, acetonitrile, DMSO, and MDM. Automated, six-way valve can add 6 types.
    Dispenser: Precision dispenser, 0.5 ml syringe, multi-capillary bundle, minimum volume 0.024 L
    Moving Probe: Fully automated X-Y axis titration module
    Cleaning: Static and dynamic cleaning stations
    Auto-Sampler: Optional, 192 sample positions, automatic mechanical gripper
    Purge Gas: 2 internal flow meters, nitrogen or argon
    Experimental Volume: 1.0-3.5 mLs
    Environmental Requirements: Indoor use, altitude ≤2000m, temperature 5℃-40℃, maximum relative humidity 80%@31℃
    Power Supply Voltage Fluctuation: +/-10%, overvoltage Cat, pollution degree 2.